AUTOLAVE |
Operation qualification test / Temperature and pressure recorder recordable real-time temperature compensation, /according to the vapor pressure built and the port Provide test report of sterile tape
Medium, glass, is used extensively as a sterilization treatment, such as a metallic test equipment or waste. / Precise and safe way to control the digital PID Jung control is possible./ Start / Stop Timer function and are made to the exhaust in a sterile setting in once./ Equipped with a level control sensor When the water level inside the chamber is not more than a certain amount / It is designed so that the product is not working to ensure the safety of sterilization subject and the experimenter. / It was produced by Compact design that can be used in a small space./ Stainless steel basket with a top-loading chamber / It indicates a temperature and for a time of two LED display at the same time. Use the touch buttons with ease, with seven images displayed can be easy to check the current progress./ The use of Stainless Steel material in the chamber and the heater, which may be generated from the high-pressure steam /Reduced risk of damage and corrosion. / Overheating of the above device as a safety device 4 to prevent the high-pressure sterilization target to protect the experimenter. / Overheating limit is attached to protect the machine and the sample. /RS-232c or 485 Interface port, software, com port cable (optional)
SPECIFICATION |
MODEL NO |
PHAS 60 |
PHAS 80 |
PHAS 100 |
Dimension_IN(WxDxH)mm |
∮345 x 650mm
|
∮395 x 660mm
|
∮445 x 670mm
|
Dimension_OUT(WxDxH)mm |
570 x 585 x 1110
|
620 x 635 x 1130
|
670 x 685 x 1130
|
Capacity |
60 Liter
|
80 Liter
|
100 Liter
|
Heater |
3kw
|
4kw
|
5kw
|
Power consumption |
14A
|
18A
|
22A
|
Plug type |
European 220 VAC STD plug
|
||
Electrical requirement |
220 VAC 50/60 Hz
|
||
Sensor |
PT 100 Ω
|
||
Temp._Controller |
Digital PID controller with timer, auto-tuning auto-sterilizer, 7 kinds of image print out |
||
Temp._Heating type |
Direct heating by stainless steel incolony heater |
||
Temp._Range |
Ambient to 131℃ |
||
Temp._Accuracy |
±1℃ at 121℃
|
||
Temp._Wait off timer |
mm:ss 59 / hh:mm 99 / Continuous selectable
|
||
Heat up time to 121 ℃ |
Approx. 20~25min
|
||
Material_in |
Stainless steel(SUS-304)
|
||
Material_out |
Steel plate with powder coating
|
||
Pressure temp. |
1.3Kg/㎤ 121℃
|
||
Pressure gags.. |
Analog type 3,0Kg/㎤ 135℃
|
||
Safety Heat |
Over heat protector, water limit sensor
|
||
Safety Pressure |
Emergency emission manual button Over pressure safety valve |
||
Safety Electrical |
Over Current & Sensor Error Detector
|
||
Pressure |
Mechanical over pressure safety valve
|
||
Basket |
2EA of Stainless steel wire basket
|
||
Option |
RS-232c or 485 Interface port, software, com port cable, thermal printer |